DETERMINING RECALIBRATION INTERVAL FOR LABORATORY EQUIPMENT

<p style="text-align: justify;">This paper describes the application of international standards, recommendations and guides, as well as other input data, to determine the interval of recalibration of laboratory equipment in the laboratory for testing the product safety testing LIND in Zenica, Bosnia and Herzegovina. An example of a pressure sensor on a 4-axis device for measuring the geometry of furniture, a detailed analysis of the recalibration interval was performed, which had to be shortened from 5 to 3 years.</p>